J Am Med Inform Assoc 14:295-303 doi:10.1197/jamia.M2219
  • Original Investigation

Assessing Data Quality in Manual Entry of Ventilator Settings

Table 5

Ventilator Management Protocol Instructions Generated with Incorrect Setting Values

Number of Protocol Instructions = 14,263
Number Incorrect Percent Incorrect
Fio2 172 1.2
PEEP 200 1.4
VT 40 0.3
Respiratory rate 66 0.5
Pressure support 116 0.8
Any incorrect 559* 3.9
  • Fio2 = fraction of inspired oxygen; PEEP = positive end expiratory pressure; VT = tidal volume; VE = minute ventilation.

  • * Some instructions were generated with more than one incorrect setting. There were 594 total incorrect settings.